1-7478-41-28 Constant Temperature Dryer (program function specification, natural convection type) 27 L with Pre-shipment Validation EOP-300V
Thông số kỹ thuật
Validated products are Azuwan original services that conduct FAT (Factory Acceptance Test) at Azuwan before shipping.
Validate (FAT) before shipping.
After receiving the order, a FAT instruction form (plan) is sent to the user, and after the signature is returned, the FAT is performed.
Appendix: FAT report/Calibration Certificate of used measuring instrument copied
*I am attaching Azwan's original FAT instruction manual (see the sample at the bottom left of the Screen).
Temperature points: 9 points at 300 degrees Celsius
Capacity (L): 27
Temperature adjustment range: room temperature + 20 ~ 300 degrees C
Maximum temperature reach time: approximately 55 minutes
Temperature distribution (JTM K05): +/- 8 degrees C (at 300 degrees C)
Temperature variation (JIS): +/- 3 degrees C (at 300 degrees C)
Temperature Gradient (JIS): 31 degrees C (at 300 degrees C)
Temperature control system: PID control, SSR output
Additional Features: Program Function Specifications/Air Jacket Natural Convection Type - > Program Function of Up to 8 Steps x 4 Patterns
Safety devices: Independent overtemperature protection device (operating temperature can be set digitally)/Overcurrent breaker/burnout mechanism (heater OFF when sensor is disconnected)
Heater capacity: 1000 W
Temperature Sensor: K thermocouples x 2 (1 each for controlling temperature and preventing excessive rise)
Power supply: 100 VAC 50/60 Hz Cord length 2 m (3P plug)
Current rating: 10.5 A
Shelf load capacity: 15 kg
Accessories: 2 shelf boards, 2 shelf rails
Local validation services (IQ/OQ) are also available. Please contact us.
*The on-site validation service is a service that performs installation qualification (IQ) and operational performance qualification (OQ) at the installation site and submits an inspection form.
*Do not put combustible materials in the cabinet.
*If you would like to request a local IQOQ, please contact us separately. We need your installation address for the estimate.